Eco2Box sets are supplied in two ways. Untreated polypropylene vessels and covers are not sterile on arrival: you autoclave them at 121 °C before first use, and we recommend using them at most 5 times. Gamma irradiated sets are supplied sterile; autoclaving them is not recommended. Which answer applies depends on the version you have, so say which one you need when you ask for samples or a quote.

What are the two ways Eco2Box sets are supplied?

ECO2 nv supplies Eco2Box vessels and filter covers either untreated or gamma irradiated. Untreated polypropylene is the route for labs that autoclave and re-use. Gamma irradiated is the route for labs that want sets supplied sterile and ready to fill.

Route On arrival Autoclave Number of uses
Untreated polypropylene Not sterile: autoclave before first use Yes, at 121 °C At most 5 recommended. The plastic takes 5 to 8, never more than 10, with clear loss of quality
Gamma irradiated Supplied sterile Not recommended: at most 2 to 3 cycles Made to be filled as supplied

Which article codes are available gamma irradiated is not published as a list. Ask us for the article you need.

How often can I autoclave an untreated Eco2Box vessel?

Autoclave untreated Eco2Box vessels and covers at 121 °C before first use. We recommend at most 5 uses. The plastic can take 5 to 8 uses and never more than 10, but its quality drops with every cycle: it becomes more brittle and breaks more easily. Container and filter keep their properties between -5 °C and +135 °C, and the filter sits in the cover, so it goes through the autoclave with the cover.

General lab guidance says the same of the material: the biosafety committee of the University of Utah describes polypropylene as a resin that can resist autoclave temperatures. ECO2 has not published an autoclave protocol (cycle time, loading, with or without medium), so run your own cycle at 121 °C, inspect the first batch, and check on your own crop that quality stays the same as the number of uses goes up.

Can a gamma irradiated Eco2Box set be autoclaved later?

We do not recommend it. Gamma irradiated Eco2Box articles are supplied sterile and are made to be filled as supplied. They take at most 2 to 3 autoclave cycles, and the plastic becomes brittle, breaks sooner and turns yellowish.

This matters for two routines. If you re-use vessels, order untreated sets. If your lab autoclaves discarded cultures inside the vessel before disposal, tell us before you order, so the route matches your waste routine. And if handling the sterile packaging is a problem in your lab, the packaging can be discussed with us.

Which route fits my lab?

  • Choose untreated polypropylene if you have autoclave capacity and want to use vessels and covers several times, at most 5 times recommended.
  • Choose gamma irradiated if you want sets supplied sterile and ready to fill.
  • Ask us if you are unsure which version you received: only gamma irradiated sets are supplied sterile.

Frequently asked questions

Are Eco2Box containers sterile when they arrive?

Only when you order them gamma irradiated. Gamma irradiated Eco2Box articles are supplied sterile. Untreated polypropylene sets are not sterile, so autoclave them at 121 °C before first use. If you are not sure which version you have, ask before you fill them.

Can the polystyrene RA trays and RDA covers be autoclaved?

No. Eco2Box RA trays and RDA covers are polystyrene and heat resistant to 85 °C, which is below the 121 °C of an autoclave. They have no filter and are not hermetic. For work that needs autoclaving, use the polypropylene sets with a filter cover.

Is there an autoclave protocol for Eco2Box vessels?

Not a published one. What ECO2 states for untreated Eco2Box polypropylene is the temperature, 121 °C, and the number of uses: at most 5 recommended, never more than 10. Cycle time and loading are not published, so run your own cycle and inspect the first batch before you scale up.

Do I need to autoclave a gamma irradiated set before use?

No. A gamma irradiated Eco2Box set is supplied sterile, so it goes straight to filling. If your protocol requires autoclaving every vessel before filling, order untreated polypropylene sets.

Request a sample set or ask a question

Tell us your crop, your stage and whether you want untreated or gamma irradiated sets, and we propose what to test. Sample sets are available on request: request a sample set or ask a question. One of the bestsellers is the O118/80 with filter, and the product range shows the other sizes.

Related guides: How many times can a polypropylene vessel be autoclaved?, Polypropylene containers for tissue culture: the facts.

Sources

  • ECO2 nv production information: untreated polypropylene autoclaved at 121 °C before first use, at most 5 uses recommended, 5 to 8 possible and never more than 10; gamma irradiated articles supplied sterile, autoclaving not recommended, at most 2 to 3 cycles; temperature range of container and filter; polystyrene trays (checked 21 November 2026).
  • University of Utah, Institutional Biosafety Committee: Autoclave use, compatible and incompatible materials (page updated 22 September 2026, read 20 November 2026).