An untreated polypropylene Eco2Box vessel and its cover are autoclaved at 121 °C before first use, and we recommend using them at most 5 times. The plastic can take 5 to 8 uses and never more than 10, but every autoclave cycle makes it more brittle, so check on your own crop that quality stays the same. Gamma irradiated sets are a separate case: autoclaving them is not recommended.

How many times can a polypropylene vessel be autoclaved?

For untreated Eco2Box vessels and covers, ECO2 nv recommends at most 5 uses, each after autoclaving at 121 °C. The plastic can take 5 to 8 uses and never more than 10, with a serious loss of quality towards the end: each cycle makes it a little more brittle, so it cracks and breaks more easily.

The cautious way to count is every autoclave run, the first one before first use included. The two general sources checked for this page give no number: the University of Utah biosafety committee says polypropylene can resist autoclave temperatures, and a CSIRO note on cultureware for algae says some plastics, like polycarbonate, can be reused and re-autoclaved “(limited times)”.

What does “more brittle” mean for a vessel in use?

Brittle means the plastic bends less before it breaks, so a part can crack where a new one would flex.

Before each re-use, look for:

  • hairline cracks or chips in the vessel or the cover
  • a cover that no longer closes onto the vessel the way a new one does
  • a part that has changed shape

Take any such vessel or cover out of use. An Eco2Box set works because the cover closes hermetically onto the vessel, so that gas exchange runs through the filter. A cracked or warped part can no longer be relied on for that. These checks are practical advice, not an ECO2 test procedure.

How long does the filter last?

ECO2 has not published a separate figure for the filter. The number of uses is given for vessel and cover, and the filter sits in the cover. What is stated: container and filter keep their properties between -5 °C and +135 °C, a range that includes the 121 °C of the autoclave.

How do I keep count in a production lab?

  • Keep vessels and covers together as a batch and log every autoclave run per batch.
  • Retire the whole batch at 5 uses, or earlier if your own crop shows a drop in quality.
  • Inspect at every run, and remove single pieces earlier when they show the signs listed above.

Frequently asked questions

Can I autoclave Eco2Box vessels more than 5 times?

The plastic can take 5 to 8 uses and never more than 10, but its quality drops with every cycle, which is why ECO2 recommends at most 5. A lab that goes beyond 5 relies on its own inspection of every vessel and cover, and on checking that its crop does as well as before.

Do Eco2Box vessels deform in the autoclave?

ECO2 has not published deformation data. Untreated Eco2Box polypropylene is autoclavable at 121 °C, and container and filter keep their properties up to +135 °C. Loading and cycle time are not published, so run a small batch in your own autoclave first and check how the covers close afterwards.

Can gamma irradiated sets be autoclaved?

We do not recommend it. Gamma irradiated Eco2Box articles are supplied sterile and are made to be filled as supplied. They take at most 2 to 3 autoclave cycles, and the plastic becomes brittle, breaks sooner and turns yellowish. If you plan to re-use, choose untreated sets before you order. If handling the sterile packaging is a problem in your lab, the packaging can be discussed with us.

Test re-use on your own autoclave

A sample set lets you run your own cycle and see how vessel, cover and filter come out. Ask for untreated sets for this test. Sample sets are available on request: request a sample set or ask a question. One of the bestsellers is the O118/80 with filter.

Related guides: Are Eco2Box containers sterile, and can I autoclave them?, Single use or reuse: the real cost of a culture vessel.

Sources

  • ECO2 nv production information: untreated polypropylene autoclaved at 121 °C before first use, at most 5 uses recommended, 5 to 8 possible and never more than 10, more brittle each cycle; gamma irradiated articles supplied sterile, autoclaving not recommended, at most 2 to 3 cycles; temperature range of container and filter (checked 21 November 2026).
  • University of Utah, Institutional Biosafety Committee: Autoclave use, compatible and incompatible materials (page updated 22 September 2026, read 20 November 2026).
  • CSIRO, Australian National Algae Culture Collection, methods: Cultureware (read 20 November 2026). The note is about algal culture, not plants.